Homogenizing System
- Update Date:2025-01-15
- Units:Instrumentation Resource Center
[YangMing] Bead Beating Grinder and Lysis System, Fastprep-24 5G

- Bead Beating Grinder and Lysis System
- Brand: MP Biomedicals
- Model: Fastprep-24 5G
- Location: Yang-Ming Campus
- Room 710-1, Biomedical Building
- Term of Use: Available anytime on demand
- First-time users must ask the center staff before use.
- Free of charge for NYCU users. Off-campus users, please contact the center before using it.
- Contact center staff:
- Extension:62382
- Email:ycirc@nycu.edu.tw
The FastPrep-24 5G is a versatile sample disruption instrument designed to quickly and efficiently homogenize, grind, or lyse biological samples. Samples are added to individual lysing matrix tubes and placed into sample holder adapters within the instrument. The unique, optimized motion of the FastPrep-24 5G shakes the samples in a tridimensional motion resulting in the collision of the sample material with the lysing matrix particles.
Specifications
- Sample Adapter: 24x 2mL QuickPrep-3 Adapter
- Speed Range:4 to 10 m/sec in 0.5 m/sec Increments; (Default: 6 m/sec for QuickPrep-3 Adapter)
- Shakes in a tridimensional motion
- Time Range:1 to 120 Seconds in 1 Second Increments (Default: 40 Seconds)
- Cycles:1 to 9 Cycles (Default: 1 Cycle)
- Pause Time During Cycles:1 to 300 Second Pause Between Cycle in 1 Second Increments (Default: 300 Seconds)
- Acceleration/Deceleration:<2 seconds to maximum speed or to stop
- Pre-Defined Assays:72 Pre-defined and Optimized Assay Programs
Major Componenets (Front View)

- This machine is on a first-come, first-use basis. First-time users should ask the center staff beforehand.
- FastPrep Manual (English version)
- FastPrep Quick Guide Manual (Chinese version)
- The samples and sample adapter must be adequately secured and balanced before running the Fastprep-24 5G.
- Press the ‘Emergency Stop Button’ and notify the center if the FastPrep-24 5G makes an abnormal sound.
- Do not exceed the maximum speed recommended by the Adapter (10 m/s).
- In case of leakage, clean the machine (do not spray alcohol directly on it) to avoid contamination of the dome and the surrounding area inside the machine; wipe with a paper towel containing distilled water and then dry it with a dry paper towel.
How to Unlock the Adapter
How to Lock the Adapter
How to Lock the Adapter